Maintenance: System Defaults - Agilent Technologies G8610A Operation Manual

Helium leak detector
Table of Contents

Advertisement

Test Port ATM
Test Port Pressure and
Spectrometer Pressure
7.4.6.2 Test Port Pressure Gauge Calibration Procedure
Calibrates the test port pressure thermocouple gauge. Perform both the Vacuum and
Atmospheric Calibration procedures in the order given.

7.4.6.3 Maintenance: System Defaults

Use this screen to reset Turbo hours plus Filaments 1and 2 hours and restore system
defaults.
This screen displays the on/time hours for the turbo pump and filaments.
When replacing a turbo pump or a filament ensure that the hours are reset. This aids in
scheduling an effective maintenance service.
Button and Fields
Restore System
Defaults
126
Agilent Technologies Helium Leak Detector
Sets the thermocouple atmospheric value. Perform this operation
only when the test port is at atmosphere.
These values appear on this screen for reference.
Figure 7-13 Maintenance: System Defaults
Resets the system back to the original factory configurations:
Auto Zero<0 - Active
Internal calibrated leak rate information will be retained
V-gain value as shipped will be retained
English Language - default
Minimum Range - E-10
Restores Setup Wizard