Agilent Technologies G8610A Operation Manual page 150

Helium leak detector
Table of Contents

Advertisement

Command
DISABLE-xREJECT
DISABLE-RANGESTOP
DISABLE-VOLUME
ENABLE-0ENABLE
ENABLE-PASSWORD
ENABLE-xREJECT
ENABLE-RANGESTOP
ENABLE-SEQUENCER
ENABLE-SPLITFLOW
ENABLE-VOLUME
EXTERNAL
INCREMENT
INTERNAL
KEEP
MANUAL
NOSNIFF
ROUGH
SNIFF
STDLEAK
TPTCATM
TPTCZERO
150
Agilent Technologies Helium Leak Detector
Table A-4 Leak Detection Actions (Continued)
Turns the Reject Set Point OFF. Where x is a number from 1 to 4 or the
letter A.
Turns the Range Stop function OFF.
Turns audio OFF.
Turns the Zero Enable function ON.
Turns the Password feature ON.
Turns the Reject Set Point ON. Where x is a number from 1 to 4 or the
letter A.
Turns the Range Stop function ON.
Turns the auto sequencer function ON.
Turns the split flow function ON.
Turns audio ON.
Sets the External Calibrated Leak (in test port) as the leak for use during
calibration.
Adds 1 to the Manual Ranging leak rate exponent (making it less negative).
Success is indicated by the normal ok response. Does nothing in
auto-ranging mode. Does nothing upon reaching the least sensitive range.
Sets the Internal Calibrated Leak as the leak to be used during calibration.
Causes the leak detector to close most valves and enter the Hold state.
Turns the Manual Ranging Mode ON. Success is indicated by the normal
ok response.
Turns the HIGH PRESSURE TEST function OFF.
Causes the leak detector to begin a roughing on the test port only.
Turns the HIGH PRESSURE TEST mode ON.
Turns Read Std Leak ON/OFF. Verifies calibration by exposing internal cal
leak to the system.
Sets the current test port thermocouple reading to represent atmospheric
pressure when the test port is exposed to atmosphere. Success is indicated
by the normal ok response.
Sets the current test port thermocouple reading to represent a ZERO
pressure reading when the test port pressure is less than 1.0E-04 Torr.
Success is indicated by the normal ok response.
Action

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G8610bG8611aG8611cG8612aG8611bG8612b ... Show all

Table of Contents