Agilent Technologies G8610A Operation Manual page 168

Helium leak detector
Table of Contents

Advertisement

Specifications
Storage Ambient
Conditions
Power Requirements
Dimensions
Weight
Touch Panel display
Regulatory
Single rotation speed
Leak test range
Maximum spectrometer
pressure
168
Agilent Technologies Helium Leak Detector
Table C-1 Leak Detector Series Specifications (Continued)
Agilent Technologies Helium Mass Spectrometer
Relative humidity: 0% to 95%, non-condensing. Temperature:
20 °C (4 °F) to +60 °C (140 °F)
Input Power
100 VAC, 50 Hz, 20 A
115 VAC, 60 Hz, 20 A or
230 VAC, 50/60 Hz, 12 A
Max. Branch Circuit Breaker: 20A with motor-rated time delay.
See outline drawings (Figure 1-1 on page 27 to Figure 1-14 on page 35).
G8610A with built in DS40M pump: 77 lb./35 kg*
G8611A w/DS-302 rotary vane pump on cart: 172 lb./78 kg*
G8611C w/IDP-15 dry scroll pump on cart: 194 lb./88 kg*
G8611B w/TS-620 dry scroll pump on cart: 192 lb./87 kg*
G8610B with IDP-3 dry pump: 77 lb./35kg*
* Add 1.0 kg if the Wireless Remote and optional I/O features are installed.
8.4 inch color TFT LCD module G084SN03 V3.
The screen format is intended to support the SVGA (800(H) x 600(V)) screen and 16.2M (RGB
8-bits) or 262k colors (RGB 6-bits). Viewing area 175.1 mm X 132.2 mm
CAN/CSA-C22.2 No. 61010-1-12 - Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use - Part 1: General Requirements
UL Std. No. 61010-1, 3rd Edition - Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use - Part 1: General Requirements
EMC Emissions and Immunity -ETSI EN 301 489-1 v2.2.0 (2017-03) Electromagnetic
Compatibility and Radio Spectrum Matters (ERM); Electromagnetic Comparability
(EMC) standard for radio equipment and services; Part1: Common technical
requirements:
ETSI EN 301489 v2.2.1 (2012-09) Electromagnetic Compatibility and Radio Spec-
trum Matters (ERM); Electromagnetic Comparability (EMC) standard for radio equip-
ment; Part 17: Broadband Data Transmission Systems
EN 61326-1: 2013 EMC requirements for Electrical equipment for measurement,
control and laboratory use - General Use
For particular Safety and EMC Standards see Declaration of Conformity.
Turbo: 70K RPM
E-11 to E-4
1E-04 Torr - Filament turns off when pressure exceeds this point
Versions