Agilent Technologies Helium Leak Detector
Table A-6 I/O Connector Pin Signals (Continued)
I/O
Signal
O
Level
Active HIGH when the test
port pressure is over set
point #4.
O
Level
Active HIGH when
calibration is successful.
I
Level
Active HIGH when Manual
Ranging is selected.
I
> 200
Active HIGH pulse.
ms Pulse
I
> 200
Active HIGH pulse.
ms Pulse
I
> 200
Active HIGH pulse.
ms Pulse
I
> 200
Active HIGH pulse.
ms Pulse
I
> 200
Active HIGH pulse.
ms Pulse
Description
Indicates that the value of
the test port pressure
exceeds the value set in the
4REJECT variable.
Indicates the calibration has
successfully completed. Set
low on power-up or if
calibration is not
successfully completed.
Activation causes the leak
detector to change from
AUTO to MANUAL ranging
; Default is AUTO.
Initiate CALIBRATING
routine with Internal
Calibrated Leak; Zeroing
routine is included.
Initiate ZEROING routine.
Initiate HOLD routine and
isolate the test port and any
test objects from leak
detector vacuum system.
Initiate VENT routine and
vent the test port to
atmosphere.
Initiate ROUGHING routine
on the test port and any
attached test objects.
Leak Detector Behavior