Section 2. Operating the Leak Detector

2.1 Screens

The leak detector display is used for the initial set-up and configuration of the leak detector.
Once the leak detector is set up and configured for a specific application, basic operation is
controlled primarily using the TEST and VENT buttons on the leak detector and the Home
screen Zero button (Section 3 "Home Screen" on page 55).
2.1.1 Basic Functions
Table 2-1 explains the icons that appear on various screens and their meaning.
Icon
Agilent Technologies Helium Leak Detector
Table 2-1 On-Screen Icons
Indicates a feature is locked and cannot be activated due to a
higher level selection in the setup.
Indicates the status of the internal calibrated leak: If flashing, the
leak calibration will expire within 10 days. Steady and blue
indicates that the leak has expired. Steady and red indicates a
temperature sensor failure or a disconnected cable to the calibrated
leak temperature sensor.
Indicates the leak detector is in split flow.
Indicates that the split flow leak rate being entered is greater than
1000 times the indicated leak rate; the maximum permitted.
Indicates that the unit is in communication with a wireless remote
unit.
Indicates that the unit has detected a fault condition.
Meaning