2. Enter the required values as shown on the calibrated leak label or calibration
certification.
3. Repeat step 1 and 2 for each desired item.
4. Press OK to exit the screen.
The value of the internal calibrated leak, as shown on the CALIBRATION Menu, is
compensated for temperature using the calibrated leak temperature sensor. The calibrated
leak is not compensated for temperature if any of the following conditions exist:
The temperature sensor on the calibrated leak fails or there is a cable connection
problem.
The temperature sensor reading is out of the operating range. The operating
temperature range is from 10.0 to 45.0° C (50 to 113° F).
The internal calibrated leak is not a E-7 or E-8 range leak.
116
Agilent Technologies Helium Leak Detector
Figure 7-5 Internal Calibrated Leak Keypads