Agilent Technologies Helium Leak Detector
7. Configure the following settings:
Bits per second (baud rate) – 9600
Data bits: – 8
Parity: – None
Stop bits: – 1
Flow Control: – None
8. Click OK.
9. Select Properties from the File menu.
The Properties window appears.
10. Configure the following settings on the Connect To tab:
Connect using – Verify the COM port is correct.
Click Configure and verify the port settings are correct.
11. Ensure that the Settings tab is configured as follows:
Terminal keys radio button selected.
Ctrl+H radio button selected.
Auto detect selected from the Emulation pull-down list.
ANSI entered in the Telnet terminal ID.
500 set in the Background buffer lines field.
12. Click ASCII Set-Up.
The ASCII Set-Up dialog box appears.
13. Ensure that the settings are as follows:
Line delay – 0 milliseconds
Character delay – 0 milliseconds
ASCII Receiving – Wrap lines that exceed terminal width
The cursor begins blinking.
14. Type some simple commands:
?LEAK
?ALL
?SETUP