I/O Connector is a 25-pin female D-sub connector to interface with an external PLC via a
parallel cable.
Inputs:
All inputs are opto-isolator LEDs with a series 3300 Ohm resistor.
A detailed summary of information related to I/O pin identifications is provided in Table A-6
on page 152.
Nitrogen Vent Port
1
2
Serial Connector
Power Entry
3
Module
4
Fore Pump Power
Connection
Analog Output
5
Chassis (Earth)
6
Terminal
50
Agilent Technologies Helium Leak Detector
Four level inputs: optically isolated, (+5 to+24) VDC.
Four momentary inputs: optically isolated, (+5 to+24) VDC, requires 200 ms
pulse width.
Outputs: Eight optically isolated emitter followers with a 10 Ohm series resistor,
14 mA max drive current, +24VDC max.
Power: An external power supply is supplied by the user.
A 9-pin male D-type connector to interface with an external PC via
RS-232 type communications protocol for system diagnostic or
control
A detailed summary of information related to serial communication
is provided in Section A "Communications Protocol" on page 139.
Provides supply power, overcurrent and EMI protection for the leak
detector. The Power Entry Module consists of:
Power Input Connector - An IEC/EN60320/C20 standard,
three-prong grounded inlet connector.
Power Switch (I/O) - A 20A rocker type Power
Switch/Thermal Circuit Breaker.
20 A EMC line filter.
An IEC/EN60320/C19 standard, three-prong grounded outlet con-
nector.
The Agilent Technologies Helium Mass Spectrometer leak detector
is equipped with a DB9F connector for the analog output. Pin #1 is
signal (positive) and Pin #2 is return (negative). Signal range: From
0 to 10 VDC
Provided for additional connection of the leak detector to the
building grounding system to improve ESD protection.