6.2.5 Password and Restrictions

The Password and Restrictions Menu is an optional configuration that protects the setup of
the leak detector. The factory and operational setups of the leak detector significantly affect
the performance of the leak detector and must only be accessed by trained service
personnel. This menu also allows protection from unintentional calibrations and accidental
operator input while the machine is unattended or under test.
Figure 6-7 Systems Settings: Password and Restrictions
Buttons and Fields
Restrictions
Set Password
Agilent Technologies Helium Leak Detector
VENT, TEST and ZERO buttons are still accessible in all
NOTE
password levels.
Activates/deactivates the password function:
Medium: Allows operation (Home Screen) of the Test, Vent
& Zero functions. Locks out all menu selections other than
the Calibration Menu. Allows Internal & External
Calibration.
High:Locks out all menu selections (Home screen only)
Changes the password. Enter up to six numeric values for the pass-
word and confirm the password selection. An empty entry is not
allowed. If no password is entered, the old password is used. The
default password from the factory is 777777. You need this when
changing to a new password.