5. Navigate to the Calibration Menu and turn the Read Standard Leak - OFF.
120
Agilent Technologies Helium Leak Detector
Adjust the ION value up or down to attain the maximum value. Ideally, the ION
value should be between 205 and 240.
Adjust the EMISSION to attain the maximum value. Ideally, the EMISSION value
should be between 0.600 and 1.200.
Set the GAIN so that the leak rate on the screen matches the value of the internal
leak you wrote down.