Table A-2 Non-Volatile Operating Parameters (Continued)
Command
INIT-ION
INIT-LANGUAGE
INIT-LINEAR
INIT-LKEXPIRE
INIT-LEAKTEMP
INIT-1LOG
INIT-ML-XFER
INIT-PASSWORD
X INIT-PWDLEVEL
INIT-QUICK-CAL
INIT-RANGESTOP
INIT-1REJECT
INIT-2REJECT
INIT-3REJECT
Agilent Technologies Helium Leak Detector
Sets the ion voltage of the ion source. Preceded by a three-digit number
(volts) in the range 220 to 280.
Selects the language to be shown on the display. Preceded by 0-6 where
(0=English, 1=French, 2=German, 3=Korean, 4=Spanish, 5=Japanese,
6=Chinese).
Sets the leak rate analog output voltage to Linear. Not preceded by a value.
Sets the date on which the internal calibrated leak expires. Preceded by the
date.
Example: mm dd yyyy INIT-LKEXPIRE
Sets the internal calibrated leak temperature (printed on the leak label)
that the factory calibration was performed at. Preceded by 12.0 to 50.0
(degrees C).
Example: 23.5 INIT-LEAKTEMP
Sets the leak rate analog output voltage to logarithmic at 1 volt per decade.
Not preceded by a value.
Sets the value of the midstage-flow mode crossover pressure in Torr.
Preceded by X.XE-XX. The range is 1.0E-03 to 1.0E-01 Torr. The default is
1.0E-01 (100 mTorr).
Sets the system password. Preceded by a 1 to 6 digit password and
followed by a digit indicating the number of digits in the password.
Example: 123123 6 INIT-PASSWORD.
Password is read back using the ?SYSPWD command.
X is the password level from 1 to 3.
Sets which type of calibration will be run if the CALIBRATE command is
initiated. Preceded by 0 or 1, 0 for FULL cal, 1 for FAST cal.
Sets the value of the most sensitive range exponent. Preceded by XX.
Sets the leak rate value for Reject Set Point #1. Preceded by a two-digit
leak rate mantissa with a decimal point after the first digit followed by E-,
and then by a two-digit leak range exponent, the helium leak rate number
is in atm cc/sec.
Example 3.4E-08 INIT-1REJECT
Same as INIT-1REJECT, but for Reject Set Point #2.
Same as INIT-1REJECT, but for Reject Set Point #3.