Maintenance: Gauge Calibration - Agilent Technologies G8610A Operation Manual

Helium leak detector
Table of Contents

Advertisement

7.4.5.1 Fore Pump Type and Exhaust Pump Activation
The fore pump type and exhaust pump are configured at the factory or can be configured by
a service engineer.

7.4.6 Maintenance: Gauge Calibration

A display of <1.0E-03 indicates the pressure is less than the lowest displayable pressure of
the TC gauge, which is 1.0E-03. Use this screen to perform a calibration on the test port
thermocouple gauge. Calibrate the thermocouple gauge for the most accurate operation.
Select Gauge Calibration from the Maintenance Menu selection.
Use this screen to:
Set a calibration point at atmosphere.
Set a calibration point at a low vacuum level (<= 1E-4 Torr).
Buttons and Fields
Test Port Vacuum
Agilent Technologies Helium Leak Detector
When the Last Service date is updated/changed the Operating
NOTE
Hours reset to Zero.
Figure 7-12 Maintenance: Gauge Calibration
Calibrates the thermocouple gauge zero value. Do this at or below
1E-4 Torr.
125