Idp-15 Shipping Locks Loosening - Agilent Technologies G8610A Operation Manual

Helium leak detector
Table of Contents

Advertisement

Agilent Technologies Helium Leak Detector
Use the provided 4 mm T-handle Allen wrench to turn the locks clockwise to
loosen; the IDP-15 ship locks only tighten/loosen, and are not removable
(Figure 1-8).
Figure 1-8 IDP-15 Shipping Locks Loosening
For the TS620 pump, use a 10 mm driver to remove the shipping bolts (one on each
side of the pump base) (Figure 1-9).
Figure 1-9 TS620 Shipping Locks Removal
31