Storage; Startup - Agilent Technologies G8610A Operation Manual

Helium leak detector
Table of Contents

Advertisement

1.3.3 Storage

If the leak detector is not used immediately, it can be stored as received without special
precautions. A dry, relatively dust-free area is preferred. The required environmental
conditions for storage are:
0% to 95% relative humidity, non-condensing
-20 °C (-4 °F) to + 60 °C (+ 140 °F) ambient temperature

1.3.4 Startup

To start the leak detector:
1. Plug the leak detector power cord into an appropriate mains supply receptacle.
2. Move the power switch on the rear power control panel to the l position. If you have a
bench system with a separate backing/roughing pump, turn the pump on. The Test
button is disabled on start up until System Ready appears on the Home screen.
Once the leak detector boots up a startup wizard appears on the touch panel display
allowing you to change the time and language.
3. Select choice of language, time/type (AM, PM & 24) and system setup.
The System Ready indication appears on the Home screen. If the leak detector is being
started after a long period of being off (several hours), it can take up to five and a half
minutes to stabilize and provide reliable, quantitative leak rate readings.
4. The leak detector is capable of performing an automated calibration routine using either
an internal calibrated leak (factory installed), or an external calibrated leak placed into
the system test port. Refer to Section 4.1 "Calibration Set-Up" on page 65 to properly
set-up the leak detector for auto-calibration.
Agilent Technologies Helium Leak Detector
Never leave the AA batteries in the Wireless Remote unit if it is
CAUTION
will not be in use for a long period of time.
37