Operation
28
S <2 ms: good
S <5 ms: acceptable
S <10ms: poor
S >10ms: not acceptable
A successful NTP Sync is achieved if the offset between
the internal clock of the device and the NTP server is
below 5ms.
The quality of the NTP synchronization can be queried
using the SCPI commands (please refer to section "Syn
chronization" on page 53 for further details).
3.5.4
Operation
The FS22 - Industrial BraggMETER SI interrogator can
be fully controlled using standard SCPI syntax com
mands (for a complete command list, please refer to sec
tion 5.1 "Communication Protocol Syntax" on page 51).
Using the Provided Drivers
These drivers can be used within the LabVIEW
ronment. To develop custom applications using this
language, copy the drivers' library supplied on the CD.
A set of ready to use examples to run on LabVIEW
also provided.
Important
Before using the provided LabVIEW
PC, make sure the Regional and Language options (in
standards and formats) are defined as dot for decimal
symbol, comma for digit grouping symbol and dash for
date separation.
A4249-8.0 HBM: public
TM
envi
TM
TM
examples on a
FS22 SI
is