HBM FS22 SI User Manual page 60

Industrial braggmeter si
Table of Contents

Advertisement

Remote Communications
60
:ACQUisition:CONFiguration:RANGe:FOR­
Mula:C:NR:[CWL;FML],...,[CWL;FML]
Defines the formula(s) and central wavelength(s) for the
range(s) on channel C. NR defines the number of ranges in
channel C, CWL represents the central wavelength of the
sensor in the specific range and FML represents the corre­
sponding formula.
Note: This command shall be used after the ranges have
been defined with the :ACQUisition:CONFigura­
tion:RANGe:WAVElength:C:NR:MIN,MAX,...,MIN,MAX com­
mand.
Note: The central wavelengths and formulas shall be sorted
by wavelength in ascending order.
Note: The variable "x" is used in the formulas to represent
the wavelength variation of the sensor relatively to the cen­
tral wavelength.
Example:
Command
Answer
:ACQUisition:CONFiguration:RANGe:FORMula:C:RN:FML
Defines the formula for the range number RN on the chan­
nel C. Range Number RN is an integer from 1 to the total
number of ranges NR.
Example:
Command
Answer
A4249-8.0 HBM: public
:ACQU:CONF:RANG:FORM:1:2:[1512.1
;-96.2*x^2+104.8*x+30],[1520.5;x*0.78]
:ACK
:ACQU:CONF:RANG:FORM:3:2:-13.2*x
^2+104.8*x+30
:ACK