BraggMONITOR SI User Interface
sensors. When performing the autoscan with this method
a pop up will appear for selecting the desired range width
(see Fig. 7.27). The minimum admissible range width is
0.5 nm while the maximum is 100 nm.
Fig. 7.27
Ranges maximum and minimum wavelengths will be set
as the detected central wavelength of the FBG (displayed
on the "Wavelengths" table as in number
12
in Fig. 7.24)
plus/minus half the range width. Maximum and minimum
range limits can then be changed on the "Ranges" table
(number
1
in Fig. 7.26).
Important
If the autoscan detects errors like overlapping ranges, it
will show the error message and will not be considered
valid. The latest successfully defined configuration will be
active.
FS22 SI
103
A4249-8.0 HBM: public