BraggMONITOR SI User Interface
Sensor Name
Central Wave
length (CWL)
Range Min. Value
(nm)
Range Max. Value
(nm)
108
Add, Edit and Delete Sensor(s)
In order to Add a sensor to a configuration, the user is
required to fill the fields (Name, Central wavelength (nm),
Range Min. Value (nm), Range Max, Value (nm) and
Formula) in the table (number
the new sensor is automatically sorted in the list by
ascendant order using its central wavelength (nm). If any
error is detected (for example, the Range Min. Value
(nm) is higher than the central wavelength (nm) an auxil
iary message is displayed in the "General Bar".
To Edit a sensor, the user can manually edit the fields
corresponding to that particular sensor. The user can
also delete a single sensor or all the sensors in an optical
channel, either selecting a sensor or the channel, respec
tively, and then pressing the "delete" button.
Sensor's identification field. It is not possible to have two
sensors with the same name.
Reference wavelength from which wavelength shift (Δλ=x) is
calculated.
Safety ranges defined to avoid crosstalk between sensors.
Each sensor has its own operation range.
It is not possible to configure sensors with overlapping
ranges. Sensor measured wavelength values out of the
defined bands are returned as -998. Inside each band only
one peak is considered.
Note: Ranges are only used/accessible when the
SPD method is active (ranges ON).
A4249-8.0 HBM: public
2
in Fig. 7.31). In the end,
FS22 SI