Single interrogator
Multiple
interrogators
FS22 SI
3.5.3
Synchronization
In order to achieve synchronous measurements between
different devices the NTP (Network Time Protocol) syn
chronization via Ethernet must be used.
Each optical Interrogator can synchronize its internal
clock with an NTP server. It is possible to achieve accu
racies of 1ms or higher, depending on whether or not a
dedicated NTP server is being used.
Further information about NTP can be found at
http://www.ntp.org.
Measuring Systems Typology
A measurement system can be as simple as a single
interrogator or a bit more complex with combined inter
rogators with the same or different sampling rates, and
interrogators combined with other equipment.
For the usage of a single interrogator no special synchro
nization is needed.
BraggMONITOR SI software can only operate one inter
rogator at a time. If an NTP server is running on the
same PC as the BraggMONITOR SI, the interrogator
internal clock – hence BraggMONITOR SI – will sync.
If more than one interrogator is used, synchronization
becomes important and if an NTP server is running the
equipment will start synching as soon as a first communi
cation from the server is received.
A4249-8.0 HBM: public
Operation
25