Operation
5 Warm up
34
:ACQUisition:SCHEdule:STARt
The scheduled acquisition start command can only be
executed in the ready state.
In this operational state the FS22 - Industrial
BraggMETER SI interrogator only replies to the com­
mands:
:ACQUisition:STOP
:IDENtification?
:STATus?
After receiving the :ACQUisition:STOP command, the
FS22 - Industrial BraggMETER SI interrogator returns to
the ready state.
In the warm up state, the FS22 - Industrial BraggMETER
SI interrogator only replies to the commands:
:IDENtification?
:STATus?
Information
The interrogator will only answer when the STATUS LED
is already blinking at 1 Hz (see section 3.3
"Switching On" on page 18).
Power Shortage
If a power shortage happens when the - FS22 Industrial
BraggMETER SI interrogator is running in the ready, free
acquisition or continuous acquisition states, the opera­
tional state is maintained once the power supply is
reestablished with the stored acquisition settings.
If the power shortage occurs in the scheduled acquisition
mode, the operational state is maintained once the power
A4249-8.0 HBM: public