Channel Power - Agilent Technologies N9913A User Manual

Fieldfox analyzers
Hide thumbs Also See for N9913A:
Table of Contents

Advertisement

Averaging
By default in ALL Channel measurements, averaging is enabled and set to
display the average of the last 15 measurements. When enabled, this average
setting is automatically making the following 'averaging' settings in order to
provide the most accurate power measurements:
 Averaging Type is set to Power.
 Trace State is set to Average.
 Detector is set to Average.
 Count is set to 15.
Any of these settings can be changed manually during a Channel measurement.
Learn more about these settings starting at page 106.
To change Averaging:
 Press Meas Setup 4.
 Then
 Then
 Enter a value between 1 and 10,000 using the numeric keypad, ▲|▼ arrows,
 Press Enter.
Traces
Only one measurement trace can be displayed in Channel Measurements.
Channel Power (CHP)
Channel Power measures total power over the specified Integrated BW. The
Integration Bandwidth (IBW) can be adjusted to measure the power over
multiple channels.
SA (Spectrum Analyzer) Mode
Average Enable ON OFF
Average Count
or the rotary knob.
Channel Power measurement; Chan 190 Downlink; GSM850 Radio Standard
OFF effectively sets the Average Count to 1.
115