How to select a Radio Standard
 Press Measure 1 OR Freq/Dist
 Then
 Then select a standard using the ▲|▼ arrows or rotary knob and press Enter.
Attenuation
The Attenuation setting controls the power level into the FieldFox.
To compare Built-in Power Meter Mode measurements to SA Channel Power
measurements, use the same attenuation settings. Built-in Power Meter Mode
defaults to 30 dB attenuation, while SA mode defaults to 10 dB attention. In
Built-in Power Meter, the attenuation setting is NOT annotated on the screen.
How to set Attenuation
 Press Scale/Amptd
 Then
 Then
 Then choose from either 30 dB or 10 dB
NOTE
The following Built-in Power Meter settings are identical to the standard Power
Meter settings. To learn about these settings, please refer to the following pages:
Set Scale ................................................................ 144
Set Relative and Offset Power ........................... 146
Set Display Units ................................................. 146
Set Averaging........................................................ 147
Single/Continuous ............................................... 147
Set Limits .............................................................. 147
Built-in Power Meter Mode - Option 310
Radio Standard
More
Internal Atten
141