Single/Continuous; Vvm Calibration; Zeroing - Agilent Technologies N9913A User Manual

Fieldfox analyzers
Hide thumbs Also See for N9913A:
Table of Contents

Advertisement

 Then
 Enter a value using the numeric keypad. Enter 1 for NO averaging.
 Press Enter.
To restart the averaging process, press Meas Setup 4, then More, then
Single or Continuous Measure
This setting determines whether the FieldFox measures continuously or only
once each time the
power or to allow you to save or analyze a specific measurement.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
 Press Meas Setup
 Then choose one of the following:
 You can also use Run / Hold +/- to toggle between Single and Continuous.

VVM Calibration

CalReady is accurate for most zeroed VVM measurements at frequencies below
about 6 GHz. For the very highest accuracy, especially above 6 GHz, then perform
a QuickCal or Mechanical Cal. Learn more on page 64.
Press

Zeroing

Zeroing, also known as Normalization, or Data/Memory, simply "Zeroes" the
Magnitude and Phase display when
measurements are relative to the measurement that was made when Zero was
pressed.
When an instrument state is saved with Zero ON, the zeroing data is saved with
the state file, similar to how a calibration is saved with a state file.
 When performing a 1-Port Cable Trimming operation, a reference cable is
 For 2-port measurements, Zero is typically pressed when a THRU connection is
Press
170
where N is the current count setting.
Avg N
Single
4
Automatically sets Continuous OFF and causes the FieldFox to
Single
o
make ONE measurement, then hold for the next Single key press. When a
data trace is displayed, the entire trace is measured, then holds. The
annotation changes to an arrow
and
Continuous OFF
o
Makes continuous measurements. This is the typical
Continuous ON
o
setting when battery power is not critical.
Zero OFF
before calibrating.
connected and measured. Then Zero is pressed, and subsequent cables are
measured and trimmed to match electrical length of the reference cable.
made in place of a DUT. Learn more on page 172.
before performing a QuickCal or Mechanical Cal.
Zero OFF
button is pressed. Use Single to conserve battery
-->
while the measurement occurs.
both put the FieldFox in
Hold
is pressed. All subsequent
Zero
Restart
Hold
Hold
mode.
FieldFox User's Guide