Trace Settings; Data Chain; Transform Enable; Gating Enable - Agilent Technologies N9913A User Manual

Fieldfox analyzers
Hide thumbs Also See for N9913A:
Table of Contents

Advertisement

With compensation for the loss using the manufacturer's specification, the
FieldFox compensates the trace as though the signal traveling through 100
meters was increased by +20 dB. Therefore the response will show 0 dB for 100%
reflection.

Data Chain

This setting, available only on the Transform Settings Table, reverts to Standard
when the FieldFox is Preset. Choose from the following:
 Standard – The normal FieldFox data processing chain. Transform
 8510 – The data processing chain used by the Agilent 8510 network analyzer.

Trace Settings

The following two settings apply to specific traces.

Transform Enable

Enable Time Domain transform for the specific trace.
 Select the trace.
 Press Measure 4
 Then
 Then choose from:
Gate Enable
Enable Gating for the specific trace.
 Select the trace.
 Press Measure 4
 Then
 Then choose from:
60
calculations are performed AFTER error correction and trace math.
Transform calculations are performed BEFORE error correction and trace
math. Learn more at the FieldFox Supplemental Online Help website:
http://na.tm.agilent.com/fieldfox/help/FieldFox.htm
Transform
On
Trace displays Time Domain data.
o
Trace displays frequency domain data.
Off
o
Transform
On
Trace displays Time Domain data.
o
Trace displays frequency domain data.
Off
o