How to set Start and Stop time
 Press Measure 4
 Then
 Then
 Then choose from:
NOTE
Zero (0) seconds is always the calibration reference plane. Negative values are
useful if moving the reference plane.

Distance Units

When markers are present on a Time Domain trace, marker X-axis values are
shown both in units of time (seconds) and distance. Select the units in which the
marker distance data is presented.
Distance Units are also used when setting Time Domain Line Loss. Learn more
on page 59.
Choose from: m (meters), or Feet.

Window Layout

This setting is used to create additional traces, or select a new layout for traces
that are already created,
Choose from: x1, x2, x2H, x3H, x4.
Learn more about multi-trace configurations on page 42.

Transform Window

There are abrupt transitions in a frequency domain measurement at the start
and stop frequencies, causing overshoot and ringing in a Time Domain response.
The Window setting reduces the abruptness of the frequency domain transitions.
This causes you to make a tradeoff in the Time Domain response.
The Window setting applies to ALL traces.
58
Transform
Transform Start Stop
Sets the transform start time.
Start
o
Stop
Sets the transform stop time.
o