Triggering - Agilent Technologies N9913A User Manual

Fieldfox analyzers
Hide thumbs Also See for N9913A:
Table of Contents

Advertisement

GSM signal, framed data format, timeslot 0 and 3 on. Sweep Time is set to
approximately the frame interval. Press Single several times until the
waveform section of interest is viewable and stable. Then markers can be
used to measure the timeslot width and interval as shown.
How to set Zero Span
 Set
How to change Sweep Time in Zero Span.
 Press Sweep 3
 Then
 Enter a value using the numeric keypad. Sweep time is limited to 1000
 Then select a multiplier key. Learn about multiplier abbreviations on page 19.
Previous FieldFox versions limited the sweep time / ResBW settings. This
restriction is no longer necessary with release A.06.25.
Learn more about SA mode Sweep Time in the Supplemental Online Help:
http://na.tm.agilent.com/fieldfox/help/FieldFox.htm

Triggering

External, Video, and RF Burst triggering allows you to initiate an SA mode sweep
using an external event such as a signal burst.
100
to the frequency of interest
Center
then set
to 0 Hz
Span
o
or select
Span. Each key press toggles the frequency span
Zero/Last
o
between 0 Hz and the last specified frequency span.
SweepTime
seconds.
FieldFox User's Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents