Frequency Range - Agilent Technologies N9913A User Manual

Fieldfox analyzers
Hide thumbs Also See for N9913A:
Table of Contents

Advertisement

Frequency Range

The X-axis frequency range determines the frequencies that are measured for
each sweep. The default Start frequency is 0 Hz. However, the Start frequency
can be set as low as –100 MHz. The internal LO of the FieldFox can be seen at 0
Hz, which will mask signals that may be present.
NOTE
Although the start frequency can be set as low as -100 MHz, amplitude accuracy
is specified above 100 kHz. Below 100 kHz, frequency accuracy is maintained,
but amplitude accuracy is degraded.
The frequency range of the measurement can be changed using the following
methods:
 Set Center and Span frequencies
 Set Start and Stop frequencies
 Toggle between Zero and Last selected span
 Set Full Span of the FieldFox
 Using Channel selection. Learn how on page 85.
 Using Marker -> Center (frequency). Learn how on page 179.
How to set Frequency Range
 Press Freq/Dist .
 Then choose from the following:
 Follow each of these by entering a value using the numeric keypad, the ▲|▼
 When using the ▲|▼ arrows, press Enter . The increment setting of the arrows
 After using the keypad, select a multiplier key. Learn about multiplier
Two more ways to change the frequency range
 Press Freq/Dist .
 Then choose from the following:
SA (Spectrum Analyzer) Mode
Center
and
Freq Span
o
entering frequency range. Enter the frequency of interest. The Center
frequency is at the exact center of the X-axis. The Frequency Span places
half of the frequency range on either side of center. When the Center and
Frequency Span values are entered, then the X-axis annotation on the screen
shows the Center and Span frequencies.
and
frequencies. Start is the beginning of the X-axis and Stop is
Start
Stop
o
the end of the X-axis. When the Start and Stop frequencies are entered, then
the X-axis annotation on the screen shows the Start and Stop frequencies.
arrows, or the rotary knob.
is based on the current span. This can be changed in SA Mode. See How to
change frequency step size below.
abbreviations on page 19.
Each key press toggles the frequency span between 0 Hz
Zero/Last Span
o
and the last specified frequency span. Learn more about Zero Span on page
99.
More
o
frequencies. This is the most common method of
83