4. Calibration Type - The FieldFox always RECOMMENDS a Cal Type that will
Begin Calibration
 Press
NOTE
If an error appears ("Failure to compute calibration steps..."), check to ensure
that the frequency range of the Cal Kit covers the frequency range of the
measurement. You can verify the frequency range of your Cal Kit at:
www.agilent.com/find/fieldfoxsupport,
 Follow the Cal Wizard prompts. Connect the specified standard at the point
 At any time, press Back to re-measure a standard if you feel it was not
 Press
CAL ON U
corrected with the Mechanical Cal.

ECal

ECal is a complete solid-state calibration solution. Every ECal module contains
electronic standards that are automatically switched into position during a
calibration. These electronic standards have been measured at the factory and
the data stored within the memory of the ECal module. The FieldFox uses this
stored data, along with the measured data, to calculate the error terms for a
measurement calibration.
You can perform the following calibrations with ECal:
 1-Port Reflection calibration
 Full 2-Port calibration
All Agilent USB ECal modules are supported. ECal modules are available in a
variety of connector types, covering many frequency ranges. For information
about available ECal modules, see
70
o
Press
Change Gender
Next...
o
Then press
For each port:
o
Use the ▲|▼ arrows or rotary knob to select the Cal Kit that you have,
o
and are going to use, for the specified ports.
Next...
o
Then press
quickly and accurately calibrate all of the displayed S-parameters. Change
from the RECOMMENDED Cal Type ONLY if you understand the
implications. Learn more about Cal Types on page 73.
To select a different Cal Type:
Press
Change Cal
Type.
o
Then using the ▲|▼ arrows or rotary knob, select a Cal Type,
o
Then press
Select and
o
Start
Calibration.
where the DUT will be connected, then press Measure.
properly connected.
to complete the calibration.
Finish
is shown on the screen for all displayed measurements that are
to change the gender of the DUT connector.
to make the selections.
to make the selection.
Finish.
Click Cal Kits.
http://www.agilent.com/find/ecal