Trace Display States

In SA Mode you can display up to four of the following types of trace states. All
SA settings are applied to all displayed traces.
A color-coded legend for displayed traces is visible in the left pane of the SA
mode screen:
W = Clear/Write; M = MaxHold; m = MinHold; A = Average; V = View
How to display Trace States
 Press Trace 6.
 Then
 Then
 Then choose from the following Trace States:
 Press
NOTE
Trace 4 data WILL be overwritten by the FieldFox when using the Independent
Source Normalize feature (page 92) or using Field Strength antenna or cable
corrections (page 89).

Average Type

In SA Mode, there are four different processes in which Averaging is performed:
 Average Traces – Learn more on page 106.
 Detection Method Averaging – Learn more on page 108.
 Noise Marker Averaging - Learn more on page 110.
 Video BW filtering (for non-Zero Span measurements) - Learn more on page
There are two types of mathematical averaging that can be performed. Select
ONE of these types and it is used for all of the above averaging processes.
106
Trace 1,2,3,4
repeatedly to select a trace number to display. Traces are
displayed in the following colors:
Trace 1 – Yellow
o
Trace 3 – Orange
o
State [current setting]
Clr /Wr
(Clear/Write) Displays the actual measured data for each sweep.
o
Displays the maximum response of the input signal at each data
MaxHold
o
point over multiple sweeps. See also Average Count on page 107.
Displays the minimum response of the input signal at each data
MinHold
o
point over multiple sweeps. See also Average Count on page 107.
Each data point is averaged over multiple sweeps as set by the
Average
o
Average Count setting. Learn more in following Average Type section.
Displays and then holds the last complete measurement trace on the
View
o
screen. This is similar to storing a trace in memory.
Displays no trace.
Blank
o
to return all traces to their default settings: Trace 1
Default All
(Clear/Write); all other traces (Blank).
96.
Trace 2 – Blue
Trace 4 – Red
.