The external battery charger (N9910X–872), lets you charge a fully depleted
battery in approximately three hours. It is a two bay, level–3 stand-alone battery
charger that is compliant with the standard Smart Battery System. The two bays
are charged sequentially on a first-come, first-served basis, though a discharge
can be performed in the left bay while the right bay is charging. It uses an
external universal switching power supply
For more information, refer to the Instruction Manual included with the external
battery charger.

Reconditioning the Battery

The battery contains electronics that monitors battery usage and tracks how
much capacity is available. This function can become less accurate because of
temperature fluctuations, aging, self-discharge, repeated partial charging, and
other factors. Reconditioning - also known as recalibrating - restores the
accuracy of the battery capacity tracking system.
Reconditioning does NOT affect actual battery performance. It only recalibrates
the charge level indicator.
Reconditioning is done by fully charging the battery, fully discharging it,
recharging it again, and then verifying that the error has been corrected.
How to determine if a battery needs reconditioning
1. Press System then
2. On the Battery screen, if the Max Error is 10% or greater, the battery needs to
 This error will affect many of the displayed battery charge indicators.
After reconditioning, if the battery is not fully charged or continues to show
more than a 9% Max Error reading, repeat the reconditioning procedure.
If the third reconditioning does not restore a full charge and give an error
reading of 9% or less, the battery may need replacement.
You can recondition a battery internally, or with the External Charger (N9910X–
872). The external charger process is faster and simpler.
How to perform Internal Reconditioning
1. Fully charge the battery inside the FieldFox (either ON or OFF) using the
2. Confirm that the battery is fully charged. When fully charged, the following
Working with the Lithium-Ion Battery
External Battery Charger LED
Green
Green flashing
Yellow flashing
Yellow/green
Red flashing
Yellow
Service Diagnostics
be reconditioned.
AC/DC adapter.
icon appears on the FieldFox:
Charging Status
Fully charged
Fast charging
Reconditioning – the accuracy of the
battery's internal LED charge gauge is
being renewed. See "Reconditioning
Batteries."
Battery is reconditioned
Error
Standby
then Battery.