Relative and Offset Power Measurements
Power measurements can be made that are relative to another reading or that are
offset by a fixed value.
Use Relative to measure the difference between power levels.
Use Offset to account for components or a long length of cable that is used
How to set Relative and Offset Power Measurements
Press Scale / Amptd .
Then choose from the following:
Display Units
Select whether to display power in dBm or Watts. When making relative power
measurements, these change to dB and %.
How to set Power Meter display units
Press Meas Setup 4 or Scale / Amptd .
Then
Then choose from the following:
Resolution
Resolution sets the number of digits to display after the decimal point on the
digital power meter display.
How to set Resolution
Press Scale / Amptd .
Then More.
Then choose from 0, 1, 2, or 3.
146
before the power sensor. For example, if using a 10 dB attenuator to protect
the FieldFox, enter an Offset Value of 10 to read the power as it would appear
without attenuation. If using a 10 dB amplifier, enter an Offset Value of –10 to
read the power as it would appear without amplification.
Relative On Off
When switched to ON, the FieldFox measures and stores
o
the current power level. Subsequent power measurements display a power
level which is relative to the stored value. When relative is ON, the units
change from dBm to dB or from Watts to %.
Toggles ON and OFF the displayed power level using the
Offset On Off
o
Offset value.
Offset Val
Value by which the power meter display is offset. A positive
o
value compensates for a component with loss. A negative value compensates
for a component with gain. Offset values are limited to +/- 100 dB.
Unit
.
Sets Power Meter display to dBm or dB.
dBm (dB)
o
Sets Power Meter display to Watts or %.
Watt (%)
o
FieldFox User's Guide