Chassis Assembly - CareFusion Alaris GP Technical & Service Manual

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Chassis assembly

Replacement Procedure
1.
Disconnect four cables from the Interface PCB and one cable from the Control PCB.
2.
Remove the two screws securing the roller mounting.
3.
Remove the roller mounting and gasket.
4.
Remove the snap rivet securing the door detector flexible circuit.
5.
Remove the two screws securing the chassis.
6.
Carefully withdraw the chassis.
7.
Reassemble in reverse order.
B
Snap rivet
Item
Description
a
Alaris GP Roller Mounting Bracket Kit
b
Alaris GP Fastener Spares Kit
A
Gasket
A
Roller mounting
bracket
1000SM00013 Iss. 4
A
Screw (x2) 90cNm
Part Number
1000SP01303
1000SP01252
60/80
Alaris® GP Volumetric Pump
Corrective Maintenance
B
Screw (x2) 65cNm

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents