Rear Case and Subassemblies

Power Supply Unit (PSU) & Speaker
Replacement Procedure
1.
Disconnect the Mains Inlet cable.
2.
Remove the three PSU screws.
3.
Remove earth wire screw and washer.
4.
Remove PSU and insulator.
5.
Pull the speaker up and out.
6.
Reassemble in reverse order.
A
PSU
Item
Description
a
Alaris GP PSU PCB Kit
b
Alaris GP Speaker KIt
c
Alaris GP Fastener Spares Kit
d*
Pad Self Adhesive Double Sided 12x12mm
* Item not shown
C
Screw/Washer (x4) 40cNm
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
Corrective Maintenance
B
Speaker & D* Self adhesive pad
Part Number
1000SP01467
1000SP01306
1000SP01252
0000ME00423
52/80