Occlusion Pressure Limits

Time to alarm following occlusion is achieved in less than 30 minutes at rates of 1 ml/h and higher by the appropriate selection of
occlusion levels.
The following graphs show the typical values for time to alarm and bolus volume that can be expected in the event of an occlusion when
the Penta Enteral 60 ml feeding syringe is selected with a Pentaferte standard administration set.
Time to alarm - 1.0 ml/h
Occlusion Level
Time to alarm - 0.1 ml/h
Occlusion Level
Tests at low alarm levels may alarm immediately - the force at these levels is commonly less than the friction in the syringe (with no
additional fluid pressure). The result is that the pressure relating to the low forces will be less than the nominal quoted occlusion
pressure.
— typical
— typical
1000DF00601 Issue 7
Alaris™ Enteral Syringe Pump
Occlusion Pressure Limits
Time to alarm - 5.0 ml/h
Occlusion Level
31/34
— typical