Near End Of Infusion (Neoi) Warning Configuration - CareFusion Alaris GP Technical & Service Manual

Volumetric pump
Hide thumbs Also See for Alaris GP:
Table of Contents

Advertisement

Near End Of Infusion (NEOI) Warning Configuration

NEOI Warning Configuration is seperate from Service Mode and can be accessed via a three-digit access code that is
entered using the following procedure:
b
1.
Hold down
and turn the pump on
2.
Enter the access code
057 using the
3.
When the code shows on screen, press
NEOI Warning
1.
Select the required NEOI status can be enabled or disbaled by pressing
OFF.
Note: ON is shown when the NEOI is disabled and OFF is shown when the NEOI
is enabled.
ACCEPT to confirm.
2.
Press
Note: ACCEPT will not be shown until after the first toggle of the status.
a
.
f
keys and the NEXT softkey.
OK to confirm.
1000SM00013 Iss. 4
Alaris® GP Volumetric Pump
ON or
NEAR END OF INFUSION
Enabled
Press ON/OFF to toggle
QUIT
14/80
Configuration and Calibration
ACCEPT
OFF