Operation
Weighing scale
Weighing process cancelled
Weighing is aborted if the patient weight (minimum 250 g)
is not removed or applied within 60 seconds.
3 short beeps are emitted.

Weighing without tare

If the last tare weight was obtained no longer than 60 minutes
earlier, and if no objects have since been removed from the
bed or placed on it, recalibration with a new tare weight is not
necessary.
Preparing for weighing
®
Set Caleo
to the horizontal position. Check with the aid of
l
the built-in water levels – see page 27.
Set both knobs to the horizontal position
l
see page 24.
Starting the weighing procedure
1
Activate the weighing process = press soft key.
®
Caleo
proposes reweighing directly without obtaining
a new tare.
2
Weigh without tare = press button.
Lay the patient on the bed.
Weighing then proceeds.
1 beep is emitted.