Advertisement

Operation
Selecting menus

Alarms

Alarm limits can be changed in the configuration (page 83).
If the skin temperature of the yellow skin temperature sensor
(skin 1) falls below the alarm limit set in the configuration:
The screen displays the warning message
»Skin 1 temperature below 36.0
The alarm tone sequence (3 beeps) is sounded,
1
The central alarm indicator lights up
2
The measured value flashes,
3
Yellow bar LED flashes.
The intermittent alarm tone sequence can be muted
for 15 minutes.
4
Suppress intermittent alarm tone = press button
or
5
Press the rotary knob.
The warning message remains on the screen,
The intermittent alarm tone is muted,
1
The central alarm indicator goes out,
2
The measured value continues to flash,
3
Yellow bar LED continues to flash.
When the measured value returns above the alarm limit:
The warning message disappears,
The intermittent alarm tone is muted,
1
The central alarm indicator goes out,
2
The measured value remains on-screen but does not flash,
3
Yellow bar LED goes out. If the extended set value range is
used the yellow bar LED will light up.
For other alarms, see "Troubleshooting – Error Messages" on
page 110.
See "Alarm description" on page 136.
*
The numerical values in this description are examples.
See "Setting alarms" on page 83.
**
The central alarm light can be switched off.
See "Setting system parameters" on page 82.