Disinfecting/Cleaning/Sterilising
Stripping down
Open the front flap
1
Turn the two locking knobs inwards as far as they will go
and fold down the front flap.
Fold out the adjustable double walls to clean them.
l
Open the side flaps in the same way.
l
Remove visible soiling with a disposable cloth and
l
detergent.
Wipe-disinfect the surfaces.
l
After allowing the disinfectant time to take effect
l
(see manufacturer's specifications), wipe the surface
with a clean damp cloth and then rub dry.
Through-hole grommets, sleeves and liners
Remove all through-hole grommets, sleeves and liners etc.
l
o
Clean them at 93
C in the cleaning and disinfecting
l
*
machine
.
Push out the drainage module
l
drainage clip.
Push the drainage clip off the drainage unit.
l
Clean the parts in the cleaning and disinfecting machine at
l
o
93
C*.
Mattress and bed
Remove the mattress from the bed.
l
2
Remove the bed.
Take care not to damage the sensor unit.
Remove visible soiling with a disposable cloth and
l
detergent.
Wipe-disinfect the surfaces.
l
After allowing the disinfectant time to take effect
l
(see manufacturer's specifications), wipe the surface
with a clean damp cloth and then rub dry.
*
Observe national and international standards on
cleaning, disinfecting and sterilisation procedures
(e.g. EN 285, EN 554, EN 556).
**