Disinfecting/Cleaning/Sterilising

Stripping down

Stripping down
Check the fresh air filter
Tilt the unit to remove the fresh air filter more easily
l
(see page 27).
1
Check the expiry date: see the label on the unit.
2
Press down the knurled part of the clip and open the
filter flap.
If a filter is already in position:
Check the condition of the filter.
l
Replace the filter when
the filter is fouled, damaged or moist.
the expiry date has been reached.
the filter has been used with an infectious patient.
Discard the old filter.
l
Fit a new filter.
l
Inserting the filter:
Insert the filter in the open filter flap.
l
2
Close the filter flap.
Check that the filter flap is securely closed.
Write down the expiry date of the new filter (2 months after
l
installation) on the label.
1
Stick the label to the device.
96
1 1