Before Reusing For A Patient - Dräger caleo Instructions For Use Manual

Neonatal incubator
Hide thumbs Also See for caleo:
Table of Contents

Advertisement

Before reusing for a patient

Check that the system has been cleaned and disinfected
l
in conformity with the regulations of the hospital.
Reassemble the equipment with disinfected hands.
l
Refit all equipment – see "Stripping down", page 96.
l
When inserting the trough, make sure that both holders
click into place.
If the holders of the trough are damaged:
Call DrägerService.
l
Fitting the water connection pipe
1
Check the O-ring (2M 50 346) for damage and replace
if necessary.
2
Check the locking valve for damage.
To make the water connection pipe easier to install, tilt the
incubator.
Connect the incubator to the mains power supply.
l
Switch on the incubator = press the on/off switch until it
l
clicks into position (see page 43).
Tilt the bed (see page 27).
l
3
Push the water connection pipe into the guide duct, holding
the LuerLock connection vertically at the top.
After overcoming a spring force,
l
4
Turn the water connection pipe counter-clockwise through
o
180
and secure the locking valve in the holder.
Disinfecting/Cleaning/Sterilising
Before reusing for a patient
2
4
1
3
103