Preparation
Mounting accessories
Monitor shelf (2M 50 085)
The monitor tray can be mounted on the wall side and/or the
handle side.
Maximum load must not exceed 20 kg!
Shelf for monitor and ventilation equipment.
This monitor shelf must only be mounted by qualified
technical staff!
To fix the monitor shelf:
— on the wall side, fit a second 38 mm pole (see page 14),
— on the handle side, fit a second base pole (see page 16)
with a 38 mm pole extension (see page 17).
Mount the shelf as specified in the Installation Instructions.
l
Fit the monitor holder = slide the shelf over both 38 mm
l
poles and
1
fix in place with the screws.
Do not exceed the maximum installation height.
Hose holder for ventilation hoses (84 11 075)
Open the front flap.
l
Raise the bed and pull it out of the incubator.
l
Push the mattress slightly to one side.
l
Place the hose holder in the mounting hole in the bed and
l
fasten from underneath with the locking screw.
Replace the bed in the incubator and close the front flap.
l
The hose holder can be fixed to any of the four corners of
the bed.
2
Clip the ventilation hoses and cables into the clips at the end
of the ventilation hose holder.
18
1
max. 20 cm
1
2
2