Checking readiness for operation
Before each use
Before using the unit, make sure that the following tests
have been performed
Disinfect hands before each test!
Check that the access ports are securely closed
1
Open the access port = press down the knurled area of the
locking bolt.
Press the access port closed until the locking bolt engages
l
in the locked position.
Try to open the access port outwards by pulling it by the
l
edge – it must not open.
If the access port does not remain securely closed:
Call DrägerService.
l
Checking the front flap and the moveable double wall for a
secure fit
Open the front flap and fold it down (see page 20).
l
2
Raise and press the flap closed and turn the two knobs
outwards until they tangibly engage in the horizontal
position.
Make sure that both knobs are engaged!
The red latches must no longer be visible!
The moveable double wall must be positioned parallel to
the front flap otherwise the warm air duct may be
obstructed!
If the front flap does not stay in place, the red latches are
visible or the double wall cannot be moved:
Call DrägerService.
l
Check that the side flap is secure
Open the side flap and fold it down (page 21).
l
3
Fold up the side flap and press it closed. Turn the two knobs
outwards until they tangibly engage in the horizontal
position.
Make sure that both knobs are engaged!
The red latches must no longer be visible.
If the side flap fails to remain engaged or if the red latches are
visible:
Call DrägerService.
l
32
1
1
2
3
3
2