Operation
Using humidity control

Alarms

In the event of water shortage
The following warning message appears on the screen
»Water empty, please refill«,
The alarm tone sequence (3 beeps) is sounded,
*
The central alarm indicator lights up
,
1
The measured value flashes,
2
Yellow bar LED flashes.
Replace the water bag or top up the water tank to the full level
marking: see page 28.
The intermittent alarm tone sequence can be muted
for 15 minutes:
3
Suppress intermittent alarm tone = press button,
or
4
Press the rotary knob.
The warning message remains on the screen,
The intermittent alarm tone is muted,
The central alarm indicator light goes out,
1
The measured value continues to flash,
2
Yellow bar LED continues to flash.
When the cause of the alarm has been remedied,
The warning message disappears,
The intermittent alarm tone is muted,
The central alarm indicator light goes out,
1
The measured value remains on the screen but does
not flash,
2
Yellow bar LED goes out.
If an extended set value range is used the yellow bar LED
will light up.
For other alarms, see "Troubleshooting – Error Messages" on
page 110.
See "Alarm description" on page 136.
*
The central alarm light can be switched off.
See "Setting system parameters" on page 82.