Dräger caleo Instructions For Use Manual page 74

Neonatal incubator
Hide thumbs Also See for caleo:
Table of Contents

Advertisement

Operation
Selecting menus
The trend graph for the selected measured value is
displayed on the screen.
Default values:
Trend 1
Skin temperature
Trend 2
Air temperature
Zoom
3 hours
The selected value is displayed as a trend graph.
In this graph, the trend curve of the measured value is overlaid
on the corresponding set value curve.
Time cursor:
The time cursor is displayed as a vertical dotted line marking
a precise point of time on the graph's time axis.
To move the time curve on the time scale = turn the
l
rotary knob.
The time marked by the cursor line is specified underneath the
dotted line by the date and time. The start time and end time of
the current time window are specified to the right and left
underneath the trend graph.
If the time cursor is moved beyond the displayed time range,
the screen adapts automatically and displays the relevant time
range:
Less recent time range = turn the rotary knob counter-
l
clockwise.
More recent time range = turn the rotary knob clockwise.
l
Data window:
The time cursor is associated with a data window situated to
the right of the displayed trend. This data window shows the
numerical values valid at the time marked on the time axis by
the time cursor.
The following values are displayed in the data window:
1
Name of the selected parameter
2
Set value of this parameter at the marked time
3
Measured value at the marked time
74
D
1
2
3

Hide quick links:

Advertisement

Table of Contents
loading