RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS Repair Instructions for PCB CIO 8601051 1. What is needed? a) Lithium battery- p/n 1835343 b) Buss wire- Kynar insulated 30awg buss wire or equivalent c) ESD Kit (Electro-static discharge kit) d) Soldering Iron (15 W w/ fine-tip) e) Solder- Kester “44”...
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS 4. Examine the area shown in Figure 2, on the PCB CIO. If there is a pull-up resistor present, remove lithium battery and continue with Step 5. If there is no pull-up resistor, no further action is necessary. Replace battery as necessary according to documented replacement interval, if necessary.
Page 5
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS 6. Briefly heat the unsoldered end of the resistor and attach one end of the buss wire. Take care to limit soldering so that the buss wire insulation does not melt and the soldered leg of the resistor does not become detached.
Page 6
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
Page 7
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS...
Page 8
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS DrägerService is a division of Draeger Medical, Inc. 3122 Commerce Drive Telford, PA 18969 Tel: (215) 721-5402 (800) 543-5047 Fax: (215) 721-5784 Web: www.draegermedical.com Printed in the U.S.A.