Operation
Using humidity control
Setting AUTO humidity
In AUTO humidity, the humidity setpoint is automatically calcu-
lated and set by the system as a function of the air temperature
setting (siehe page 133).
The maximum relative humidity depends on the ambient
temperature and incubator air temperature at max. 75 %
relative humidity.
1
Switch humidity control over to AUTO humidity =
press button (soft key).
2
Activate AUTO humidity = press the rotary knob.
–
The display returns to the standard screen. The measured
value and the automatic set value are displayed.
Manually adjusting the set value
Standard set value range
3
Switch humidity control over to manual mode = press button
(soft key).
4
Increase set value = turn rotary knob clockwise.
4
Decrease set value = turn rotary knob counter-clockwise.
4
Confirm set value = press rotary knob.
The maximum attainable humidity level depends on the
air temperature and ambient humidity.
At high air temperatures or low ambient humidity, the attainable
®
humidity level in Caleo
is reduced.
58
30 % to 99 %
D
1
D
3
2
4