Dräger caleo Instructions For Use Manual page 24

Neonatal incubator
Hide thumbs Also See for caleo:
Table of Contents

Advertisement

Preparation
Doors, ports and bed
Using the X-ray drawer
The X-ray drawer can be pulled out when the front flap is either
open or closed.
To open the X-ray drawer:
1
Turn both knobs to the horizontal position marked
2
pull the drawer out by the recessed handle or the knobs.
Insert or remove the X-ray cassette. Recesses are provided
l
in the X-ray drawer for positioning.
Do not use the X-ray drawer as a writing support or bed for
the patient.
To close the X-ray drawer:
2
push the drawer inwards under the bed until it tangibly clicks
into place.
Push the X-ray drawer in fully! Otherwise the hot air duct
will be interrupted, and the patient may be warmed or
cooled excessively!
Sealed through-holes
3
Tubing ports (2M 50 412)
4
Tubing grommets (2M 50 385)
5
The feeding grommet plug (2M 51 109) can only be used if
a double wall is not attached to the canopy.
Route the cables or hoses through the flexible grommets
l
and sleeves.
To route ventilation hoses and cables through the Caleo
use the ventilation hose holder (page 18).
*
Hose fixing devices
The hose fixing devices (2M 51 140) can only be attached to
hose modules with pins (2M 51 139). Hose fixing devices act
as guides for hoses and cables.
*
optional equipment feature