Disinfecting/Cleaning/Sterilising
Before reusing for a patient
Assembling the double wall
Check parts for damage and replace if necessary.
l
Fit the parts into the double wall.
l
Fit in the following order:
1
Sleeve
2
Plunger
3
Centering lug
4
Seal
Inserting the drainage module
Check parts for damage and replace if necessary.
l
Push the drainage unit and the drainage clip back together.
l
Make sure that they are aligned correctly!
5
Insert the drainage module into the pillar element from the
side until it engages.
The lug of the drainage clip must be facing outwards and the
perimeter of the drainage module must be flush with the
pillar element.
Close the side flap.
l
Check that the incubator is ready for operation –
l
see "Before each use" on page 31.
To remove possible disinfectant residue, it is recommend
running the incubator in standby mode.
®
Switch on Caleo
(see page 43).
l
Activate air temperature control (see page 44).
l
o
®
Run Caleo
at 37
C with opened flaps.
l
If using a water container:
Do not refill the water container until just before placing the
patient in the incubator (see page 28)!
If using a water bag:
Do not connect the water bag until just before placing the
patient in the incubator (see page 28)!
*