Dräger caleo Instructions For Use Manual page 84

Neonatal incubator
Hide thumbs Also See for caleo:
Table of Contents

Advertisement

Operation
Configuration
Initial volume (sound intensity) of the alarms:
Adjustment range:
1 to 8
Default:
1
The initial sound intensity is highlighted by a light
background.
Set an initial sound intensity = turn rotary knob.
l
Confirm initial sound intensity = press rotary knob.
l
The screen returns to the alarm settings menu (page 83).
Cancel selection = press »
l
The display returns to the configuration parameter menu
(see page 80).
Alarm limits:
Adjustment ranges:
Air temperature deviation
Skin temperature deviation
O
deviation*
2
Default values:
Air temperature deviation
Skin temperature deviation
O
deviation*
2
Select the alarm limit for air/skin temperature and O
l
turn and push the rotary rotary knob.
The alarm limit setting for the deviation in air temperature is
highlighted by a bright border.
Select air temperature deviation = turn the rotary knob.
l
Confirm air temperature deviation = press the rotary knob.
l
The screen returns to the alarm settings menu (see page 83).
The other alarm limits are set in the same way.
Cancel selection = press »
l
The display returns to the configuration parameter menu
(see page 80).
*
optional
84
« soft key button.
o
–1.5 or –2.5
C
o
±0.3 to 1.0
C
±3 % or ±5 %
o
–1.5
C
o
±0.5
C
±5 %
*
=
2
« soft key button.