Checking readiness for operation
Before each use
1
Grasp the handles on the sides of the canopy with both
hands.
2
Lift the canopy horizontally off the pillar elements.
If the canopy holders are damaged:
Call DrägerService.
l
Check that the trough is secure
Remove the canopy.
l
Remove the mattress.
l
Remove the bed.
l
3
Check the catches of the trough.
Place the bed on the trough.
l
Place the mattress on the bed.
l
Refit the canopy.
l
If the catches for the trough are damaged:
Call DrägerService.
l
Remove the bed
Open the front flap and fold it down.
l
4
Turn both knobs to the vertical position marked
5
Grasp the bed by the recessed handle or by the knobs and
pull it out towards the front as far as it will go.
5
Push the bed back until it clicks into place,
4
Turn both knobs to the position marked
Close the front flap.
l
If the bed cannot be pulled out or pushed in or if the knobs are
damaged:
Call DrägerService.
l
If the bed is not fully pushed in, the hot air duct will be
interrupted, causing the control system to malfunction!
The result may be excessive cooling or overheating of
the patient!
34
1
3
.
.
2
5
4
1
2
3
4